Dissolved Ozone Controller / Transmitter
This innovative device is expertly designed for seamless online measurement and precision control of water disinfection across diverse areas such as the tap water industry, swimming pools, and water parks. It's an indispensable tool for ozone generators and container disinfection in the food and beverage industry. However, please note that it is not suited for ozone disinfection produced by electrolytic water generators.
Boasting a 128 x 64 dot-matrix LCD display, this analyzer offers switchable languages in both Chinese and English. With an IP65 protection level, it ensures stable, reliable performance in any weather condition.
The user-friendly design features a simple menu layout for effortless operation. With graphical prompts and a clear, aesthetically pleasing interface, it enhances user experience.
Equipped with adjustable software digital filtering and advanced hardware resistance to interference, this device ensures more stable and accurate measurements, seamlessly adapting to complex industrial environments.
Offers a versatile global access power supply range from 85~260VAC, with an option for a customized DC model ranging from 18 to 36 V D C, catering to diverse power requirements.
Incorporating an RS-485 digital interface and the MODBUS-RTU communication protocol, this device supports bi-directional communication, enabling complete remote control of the instrument with ease.
Waterworks |
Drinking water |
Swimming pool |
Food hygiene |
ozone |
measuring range |
0.00~20mg/L.00 |
resolution ratio |
0.01 mg/L |
Measurement accuracy (electronic unit) |
±0.10 mg/L |
temperature |
measuring range |
-10.0~130.0ºC |
resolution ratio |
0.1ºC |
Measurement accuracy (electronic unit) |
±0.3ºC |
Temperature input |
PT1000 |
temperature compensation |
Automatic / manual |
Transfer current |
Output, type |
Two roads of 4~20 mA (the corresponding range can be set) |
Current accuracy |
±1% F .S |
output loading |
less-than 500Ω |
control |
Functional relay |
One (which can be set to the cleaning or alarm function) |
Switch amount relay |
2 SPST relays |
load capacity |
2.5A 230VAC |
data transmission |
interface |
All-way RS485 isolation voltage is 2500Vrms |
protocol |
MODBUS-RTU (read-write two-way communication) |
Other parameters |
working power supply |
85~260VAC or 18~36VDC (optional before order) |
working temperature |
0~60ºC |
Operating humidity |
Relative humidity was <90% |
levels of protection |
IP65 |
way to install |
Disk installation |
outline dimension |
(H×W×D) 108×108×132 mm |
Open hole size |
92.5 * 92.5 mm (positive tolerance) |